Waters 2489 UV / Visible Detector Dual Wavelength with 10mm Flow Cell ( 1000 PSI / 70 Bar Max Pressure ) for HPLC Systems
Working. The unit has been tested and powers up; no errors were displayed, and the lamp option turned green, indicating ready to use. The D2 lamp is in working condition. Please refer to the pictures.
Performance Specifications:
Wavelength range: 190 to 700 nm.
Bandwidth: ≤5 nm.
Wavelength accuracy: ±1.0 nm (via patented erbium filter).
Wavelength repeatability: ±0.1 nm.
Optical Component & Flow Cell Specifications:
Light source: Deuterium arc lamp. Warranty: 2000 hours or 1 year (whichever comes first)
Photodiodes: 2 silicon photodiodes (matched pair)
Flow cell design: Patented TaperSlit
Cell volume: 16.3 µL (Standard analytical cell)
Path length: 10 mm (Standard analytical cell)
Pressure limit: 6895 kPa (69 bar, 1000 psi)
Used
Working. The unit has been tested, and powers up; no errors were displayed. The D2 lamp is in working condition. Please refer to the pictures.
33lbs
2489
United States Shipping
This product qualifies for free shipping to the continental U.S.
International Shipping
We ship internationally. Buyers are responsible for any import duties, taxes, or customs requirements. Contact us to request an invoice for customs clearance. We recommend checking with your local customs authority before purchasing.
Overnight Shipping
We are happy to accommodate expedited shipping using your UPS or FedEx account. Please contact us to provide your account information.
This product qualifies for a free return.
We have a 30-day return policy from the date of receipt.
To learn more please visit our Return Policy.